InstrumentsPressure and temperature measurementT +49 40 5247 4156 · M +49 176 3130 5711sales@eurekaeu.com
Pressure gauges with bar scales on insulated refrigeration lines next to a compressor

Home / Applications / Refrigeration

Industry · Refrigeration

Instruments for refrigeration plants, compressors and ammonia circuits

Pressure gauges graded in pressure and refrigerant temperature, duplex and differential gauges for compressors, pressure transmitters, thermometers and temperature sensors. Specified to your refrigerant, range and connection – and quoted by us.

  • Refrigerant scales incl. R717 (ammonia)
  • Duplex and differential gauges
  • Liquid-filled, vibration-protected
  • More than 40 connection threads
−1 … +25 barstandard refrigeration ranges
16refrigerant scales on the datasheet
Class 1.0accuracy (1 % FS)
DN 63 – 160dial sizes

Requirements

What a refrigeration plant asks of an instrument

Refrigerant, compressor vibration and the plant layout decide the instrument. These are the points we check with you before we quote – all values from the manufacturer’s datasheets.

01

Refrigerants and scales

  • Refrigeration scales graded in pressure and temperature
  • Scales for R22, R134a, R236fa, R290, R404, R407, R410, R422, R427, R449, R502, R507, R717, R744, R1234ze and R1270 – other media on request
  • Measuring system in brass/bronze, steel or AISI 316 to suit the refrigerant
02

Ranges and accuracy

  • Standard ranges −1…0…+12 bar and −1…0…+25 bar; other ranges on request
  • Accuracy class 1.0 (1 % FS)
  • Dial sizes DN 63, 80, 100 and 160
03

Compressors and oil filters

  • Duplex gauges show two pressures on one dial, DN 80, 100 and 160
  • Differential gauge intended for compressors, measuring suction and oil pressure
  • Also used for the differential pressure over oil filters
04

Vibration and mounting

  • Liquid filled, vibration protected and long-term stable
  • Bottom or rear connection; direct, wall or panel mounting
  • More than 40 connection threads from a modular nipple programme
05

Signals and temperature

  • Pressure transmitters for direct mounting
  • Resistance thermometers and vapour-pressure thermometers
  • Mechatronic thermometer for temperatures up to 300 °C
✓

Documentation

Tell us the refrigerant, the plant standard and the documents you need. We confirm per model what can be supplied and state it in the quotation.

In service

Where refrigeration instruments work

Pressure gauge on a tapping tee between insulated refrigeration lines
Gauges on insulated suction lines
Gauge panel with three pressure gauges on a refrigeration unit
Compressor pressure gauges
Refrigeration compressor with suction and discharge pressure gauges
Industrial reciprocating compressor
Reciprocating ammonia compressors with local pressure gauges in an industrial refrigeration plant
Ammonia compressors with local gauges
Row of pressure gauges with needle valves on a refrigeration header
Row of local gauges

Photos show typical installations. Image credits.

Typical applications

How instruments are typically used in refrigeration

Two common scenarios. They describe typical installations, not a specific plant or customer.

Gauge panel with three pressure gauges on a refrigeration unit
Scenario 1

Suction and oil pressure on a compressor

The difference between suction and oil pressure shows whether the compressor is lubricated properly.

  1. A differential refrigeration gauge is mounted on the compressor and shows suction and oil pressure.
  2. Where suction and discharge are read side by side, a duplex gauge shows both pressures on one dial.
  3. The scale is graded in pressure and in the temperature of the refrigerant in use.

What we need from you: refrigerant, ranges, connection thread, mounting and quantity.

Pressure gauge on a tapping tee between insulated refrigeration lines
Scenario 2

Ammonia (R717) circuit in a machine room

In an ammonia plant, each line needs a local reading and the control system needs the same values as a signal.

  1. Liquid-filled refrigeration gauges with an R717 scale give the local reading on each line.
  2. Pressure transmitters for direct mounting feed the plant controller.
  3. Resistance thermometers and dial thermometers cover the temperature points.

What we need from you: refrigerant, ranges, wetted material, signal type and number of measuring points.

Engineering

Technical notes and common faults

General engineering points we check with you. They are not a guarantee for a specific installation.

Ammonia and copper alloys

Ammonia attacks copper alloys. For R717 the measuring system and connection are specified in steel or AISI 316.

Wrong refrigerant scale

A temperature scale only fits the refrigerant it is graded for. Name the refrigerant on the enquiry.

Icing and condensation

Cold lines collect moisture on and in the case. Mounting position and case filling are agreed before the quotation.

Documents

Datasheet with the quotation. Calibration documents and, where available for the product type, certificates are confirmed per model. Resources and checklists →

Why FEX

What you get with us

Calibrated one by one

Each pressure gauge is calibrated manually and individually at five points on the scale, upwards and downwards. Bourdon tubes are heat-treated to reduce hysteresis.

Built to your specification

Standard products and customer-specified versions come from one modular system – ranges, threads, fills and mounting to suit the measuring point.

Documents with the order

Datasheet with the quotation. Calibration documents and, where available for the product type, certificates are prepared to match your order.

One contact in Hamburg

Selection, quotation, documentation, shipping to your site and after-sales – coordinated from Hamburg.

Send us your refrigeration specification

  1. Refrigerant
  2. Range and unit
  3. Connection and mounting
  4. Single, duplex or differential
  5. Output: dial, signal or contacts
  6. Quantity and delivery country

Ready to specify an instrument?

Send us the measuring point and requirements – we reply with a written quotation and the matching datasheet.

FEX Global · Am Kaiserkai 13, 20457 Hamburg · T +49 40 5247 4156 · M +49 176 3130 5711 · sales@eurekaeu.com